Ten-year socket wear in 66 hip arthroplasties: Ceramic versus metal heads

  • Hans M. Schüller
  • , Rene K. Marti
  • University of Amsterdam

Research output: Contribution to journalArticleAcademicpeer-review

Abstract

Radiographic measurements of polyethylene socket wear were performed in 33 hip prostheses with an Al2O3-ceramic head and 33 prostheses with a metal head. At 9-11 years, the average reduction of the diameter of the polyethylene was 0.26 and 0.96 mm, respectively, in the ceramic and the metal group. However, the number of clinically proven or radiographically suspected loosenings did not differ between the two groups. © 1990 Informa UK Ltd All rights reserved: reproduction in whole or part not permitted.
Original languageEnglish
Pages (from-to)240-243
JournalActa orthopaedica
Volume61
Issue number3
DOIs
Publication statusPublished - 1990