Skip to main navigation Skip to search Skip to main content

Technical problems concerning the use of immunoblots for the detection of antinuclear antibodies

Research output: Contribution to journalArticleAcademicpeer-review

Abstract

When the immunoblotting technique is used as a diagnostic tool, the reproducibility of the results is a major problem. When purified radiolabelled proteins were applied onto SDS gels, the recovery of radioactivity on the blot after electrophoresis, blotting and incubation ranged from 10 to 65%, depending on the protein. Although the addition of SDS was subsequently shown to improve protein transfer from gel to blot, it is not recommended because immunological recognition of proteins is diminished after this transfer step. We suggest that during the incubation of protein blots detergents are necessary not only to diminish non-specific background, but also to renature proteins. However, since these detergents also elute protein from nitrocellulose and other blotting matrices, they are in part responsible for the lack of reproducibility in immunoblotting results. © 1990.
Original languageEnglish
Pages (from-to)63-70
JournalJournal of immunological methods
Volume129
Issue number1
DOIs
Publication statusPublished - 1990

Fingerprint

Dive into the research topics of 'Technical problems concerning the use of immunoblots for the detection of antinuclear antibodies'. Together they form a unique fingerprint.

Cite this