Six-lead device superior to single-lead smartwatch ECG in atrial fibrillation detection

  • Amsterdam UMC - University of Amsterdam
  • Department of Cardiology, Tergooi Ziekenhuis, Blaricum, The Netherlands

Research output: Contribution to journal › Comment/Letter to the editor › Academic

60 Downloads (Pure)

Abstract

This was a head-to-head comparative study on different electrocardiogram (ECG)-based smartwatches and devices for atrial fibrillation detection. We prospectively included 220 patients scheduled for electrical cardioversion and recorded ECGs with 3 different devices (Withings Move ECG, Apple Watch 5, Kardia Mobile 6-leads) as well as the standard 12-lead ECG (gold standard), both before and after cardioversion. All atrial fibrillation detection algorithms had high accuracy (sensitivity and specificity: 91–99%) but were hampered by uninterpretable recordings (20–24%). In cardiologists’ interpretation, the 6-lead device was superior (sensitivity 99%, specificity 97%) to both single-lead smartwatches (P < .05) for atrial fibrillation detection.

Original languageEnglish
Pages (from-to)53-58
Number of pages6
JournalAmerican heart journal
Volume253
DOIs
Publication statusPublished - 1 Nov 2022