Performance and safety of the PRICO closed-loop oxygen saturation targeting system in neonates: pragmatic multicentre cross-over study (TarOx Study)

  • National Institute of Medical Statistics
  • Czech Technical University in Prague
  • University of Medical Sciences Poznan
  • Pomeranian Medical University in Szczecin
  • Department of Neonatology
  • Ujastek Medical Center

Research output: Contribution to journalArticleAcademicpeer-review

55 Downloads (Pure)

Abstract

OBJECTIVE: This study aims to evaluate the performance of the fabian-Predictive-Intelligent-Control-of-Oxygenation (PRICO) system for automated control of the fraction of inspired oxygen (FiO2).

DESIGN: Multicentre randomised cross-over study.

SETTING: Five neonatal intensive care units experienced with automated control of FiO2 and the fabian ventilator.

PATIENTS: 39 infants: median gestational age of 27 weeks (IQR: 26-30), postnatal age 7 days (IQR: 2-17), weight 1120 g (IQR: 915-1588), FiO2 0.32 (IQR: 0.22-0.43) receiving both non-invasive (27) and invasive (12) respiratory support.

INTERVENTION: Randomised sequential 24-hour periods of automated and manual FiO2 control.

MAIN OUTCOME MEASURES: Proportion (%) of time in normoxaemia (90%-95% with FiO2>0.21 and 90%-100% when FiO2=0.21) was the primary endpoint. Secondary endpoints were severe hypoxaemia (<80%) and severe hyperoxaemia (>98% with FiO2>0.21) and prevalence of episodes ≥60 s at these two SpO2 extremes.

RESULTS: During automated control, subjects spent more time in normoxaemia (74%±22% vs 51%±22%, p<0.001) with less time above and below (<90% (9%±8% vs 12%±11%, p<0.001) and >95% with FiO2>0.21 (16%±19% vs 35%±24%) p<0.001). They spent less time in severe hyperoxaemia (1% (0%-3.5%) vs 5% (1%-10%), p<0.001) but exposure to severe hypoxaemia was low in both arms and not different. The differences in prolonged episodes of SpO2 were consistent with the times at extremes.

CONCLUSIONS: This study demonstrates the ability of the PRICO automated oxygen control algorithm to improve the maintenance of SpO2 in normoxaemia and to avoid hyperoxaemia without increasing hypoxaemia.

Original languageEnglish
JournalBMJ Paediatrics Open
Volume8
Issue number1
DOIs
Publication statusPublished - 16 Jul 2024

Keywords

  • Humans
  • Cross-Over Studies
  • Infant, Newborn
  • Oxygen Saturation
  • Female
  • Male
  • Intensive Care Units, Neonatal
  • Hypoxia
  • Hyperoxia/prevention & control
  • Oxygen/blood
  • Oximetry/methods
  • Oxygen Inhalation Therapy/methods
  • Respiration, Artificial/adverse effects
  • Infant, Premature

Fingerprint

Dive into the research topics of 'Performance and safety of the PRICO closed-loop oxygen saturation targeting system in neonates: pragmatic multicentre cross-over study (TarOx Study)'. Together they form a unique fingerprint.

Cite this