Abstract
The calibration of a chemical sensor for chlorinated hydrocarbon analytes based on the Fujiwara reaction is described. This sensor generates a particular type of data: medium-rank second-order data. With this type of data it is possible to calibrate the sensor in such a way that quantitation for the analytes in the presence of unknown interferents is possible. The calibration method developed is a new approach based on so-called restricted Tucker models that utilize all available chemical information
| Original language | English |
|---|---|
| Pages (from-to) | 3345-3351 |
| Journal | Analytical chemistry |
| Volume | 66 |
| Issue number | 20 |
| DOIs | |
| Publication status | Published - 1994 |
Fingerprint
Dive into the research topics of 'MULTICOMPONENT DETERMINATION OF CHLORINATED HYDROCARBONS USING A REACTION-BASED CHEMICAL SENSOR .3. MEDIUM-RANK 2ND-ORDER CALIBRATION WITH RESTRICTED TUCKER MODELS'. Together they form a unique fingerprint.Cite this