  • M. van Herk
  • , A. Bel
  • , K. G. Gilhuijs
  • , J. V. Lebesque
  • , A. van Dalen
  • , P. van der Ven
  • , R. Vijlbrief

Research output: Contribution to journalArticleAcademicpeer-review

Abstract

In our institute, we have developed an electronic portal imaging system based on a matrix of 256 x 256 ionisation chambers. By improvements to the electronics, the system produces images with the same quality as the original system but 3-10 times faster. Software for automatic image analysis has been applied to more than 10,000 images over the last two years. Using an off-line correction strategy, the systematic patient set-up error has been limited to 5 mm or less for 98% of the patients treated for prostate cancer
Original languageEnglish
Pages (from-to)601s-606s
JournalBulletin du cancer
Volume82
Issue numberSuppl. 5
Publication statusPublished - 1995