Skip to main navigation Skip to search Skip to main content

Clinical feasibility of monitoring m-THPC mediated photodynamic therapy by means of fluorescence differential path-length spectroscopy

  • Baris Karakullukcu
  • , Stephen Chad Kanick
  • , Jan Bonne Aans
  • , Henricus J. C. M. Sterenborg
  • , I. Bing Tan
  • , Arjen Amelink
  • , Dominic J. Robinson

Research output: Contribution to journalArticleAcademicpeer-review

Abstract

The objective quantitative monitoring of light, oxygen, and photosensitizer is challenging in clinical photodynamic therapy settings. We have previously developed fluorescence differential path-length spectroscopy (FDPS), a technique that utilizes reflectance spectroscopy to monitor microvascular oxygen saturation, blood volume fraction, and vessel diameter, and fluorescence spectroscopy to monitor photosensitizer concentration. In this paper the clinical feasibility of the technique is tested on eight healthy volunteers and on three patients undergoing PDT of oral cavity cancers. Model-based analysis of the measured spectra provide quantitative tissue parameters that are corrected for background tissue absorption, autofluorescence, and the transmission of the optical system; this method allows comparison of intra- and inter-subject parameters. The FDPS correctly estimated the absence of m-THPC in volunteers and detected photobleaching in the areas receiving treatment light in patients undergoing PDT treatment. This study demonstrates the feasibility of monitoring clinical photodynamic therapy treatments using optical spectroscopy
Original languageEnglish
Pages (from-to)740-751
JournalJournal of biophotonics
Volume4
Issue number10
DOIs
Publication statusPublished - 2011

UN SDGs

This output contributes to the following UN Sustainable Development Goals (SDGs)

  1. SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well-being

Fingerprint

Dive into the research topics of 'Clinical feasibility of monitoring m-THPC mediated photodynamic therapy by means of fluorescence differential path-length spectroscopy'. Together they form a unique fingerprint.

Cite this