Abstract
Intubating introducers are often used to facilitate difficult endotracheal intubation. We report a case in which a Frova catheter was used for insertion of a laser-resistant endotracheal tube and in which a strip of plastic was shaved off from the catheter. The plastic strip formed a convolute mass, causing relevant airway obstruction. Clinicians should be aware of this potentially dangerous complication and should be cautious in using Frova intubating introducers in combination with laser-resistant tubes.
| Original language | English |
|---|---|
| Pages (from-to) | 59-61 |
| Journal | Journal of clinical anesthesia |
| Volume | 32 |
| DOIs | |
| Publication status | Published - 1 Aug 2016 |
Keywords
- Intubation
- Intratracheal
- Foreign bodies
- Airway obstruction
- Equipment failure
- Intraoperative complications