A comparative study of the Osstell™ versus the Osstell Mentor™ to evaluate implant stability in human cadaver mandibles

R. S. Snijders, A. J. van Wijk, J. A. Lindeboom

Research output: Contribution to journalArticleAcademicpeer-review

7 Citations (Scopus)

Abstract

The aim of this study was to assess the primary stability of dental implants by resonance frequency analysis (RFA) using the Osstell™ and Osstell Mentor™ devices and to investigate the reproducibility and comparability of the results obtained with these devices. Twenty-four Straumann implants (Straumann AG, Basel, Switzerland) were placed in the anterior mandible of 12 fresh edentulous human cadaver mandibles. The implant stability quotients (ISQs) were measured with the Osstell™ and Osstell Mentor™ when implants were inserted at 50% of their length and following their complete insertion. The Osstell™ device measured lower scores compared with the Osstell Mentor™. This was significant for the full position (mean difference = 9·9), t (11) = 7·4, P < 0·001 and for the halfway position (mean difference = 5·9), t (11) = 2·41, P = 0·03. In conclusion, the Osstell™ produced relatively lower ISQ scores than the Osstell Mentor™
Original languageEnglish
Pages (from-to)774-779
JournalJournal of oral rehabilitation
Volume40
Issue number10
DOIs
Publication statusPublished - 2013